Thermodynamic response of an ultrasonic densitometer
Fecha
2005-12-01Registro en:
International Congress on Noise Control Engineering 2005, INTERNOISE 2005, v. 5, p. 4137-4144.
2-s2.0-84873818773
6405339510883203
2883440351895167
0000-0003-4201-5617
0000-0001-6320-755X
Autor
Universidade Estadual Paulista (Unesp)
Universidade de São Paulo (USP)
Institución
Resumen
An ultrasonic densitometer is experimentally characterized as a function of temperature. The measurement of propagation velocity and density of several liquids is performed in the 15 to 40°C temperature range. Results are compared to tabulated values in the case of distilled water, showing accuracy of 0.07% for the propagation velocity. Density values are compared to those obtained with a pycnometer, showing 0.2% accuracy in density measurement for stabilized temperature and 0.4% under thermal gradient conditions.