Artículos de revistas
Reversible intermittent flow-injection determination of mercury in sediments and vinasses by cold vapor atomic absorption spectrometry
Fecha
1999-01-01Registro en:
Laboratory Robotics and Automation. New York: John Wiley & Sons Inc., v. 11, n. 5, p. 304-310, 1999.
0895-7533
10.1002/(SICI)1098-2728(1999)11:5<304
WOS:000083127200010
4148837403618207
9165109840414837
Autor
Universidade Estadual Paulista (Unesp)
Institución
Resumen
A reversible intermittent pow-injection procedure is proposed for the automated determination of mercury in sediments and vinasses by cold vapor atomic absorption spectrometry, CVAAS. Solutions of sample and stannous chloride are carried by two air streams and sequentially injected into the generator/separator chamber in a segmented asynchronous merging zone configuration. The intermittent flow in the forward direction carries the mercury vapor through the flow cell, and in the backward direction, if aspirates the the remaining solution from the vessel to waste. We investigated composition and concentration of reagents, pow rates, commutation times, reactor configuration, and conditions for mercury release. The accuracy was checked by mercury determination in a certified sediment and spiked vinasses and river waters. The system handles about 100 samples per hour (0.50-5.00 mu g L-1), consuming ca. 2.5 mL of sample and 50 mg of SnCl2 per determination; Good recoveries (92-103%) were obtained with spiked samples. Results are precise (RSD <3% for 2.5 mu g Hg L-1, n = 12) and in agreement with values for certified reference material at 95% confidence level. (C) 1999 John Wiley & Sons, Inc.