Artículos de revistas
Cathodic stripping voltammetric determination of cefaclor in pharmaceutical formulations
Fecha
1999-01-01Registro en:
Analytical Letters. New York: Marcel Dekker Inc., v. 32, n. 1, p. 97-109, 1999.
0003-2719
10.1080/00032719908542601
WOS:000078552100008
0000-0002-2296-1393
Autor
Universidade Estadual Paulista (Unesp)
Univ Loughborough
Institución
Resumen
A sensitive method is described for the determination of cefaclor by cathodic stripping voltammetry at the hanging mercury drop electrode. cefaclor is accumulated at the electrode surface as a mercury salt, which is reduced at -0.67 V. The optimum accumulation potential and accumulation time were +0.15 V and up to 180 s, respectively. Linear calibration graphs were obtained between 3.9 mu g.L-1 to 39 mu g.L-1 and the limit of determination was evaluated to be 1.9 mu g.L-1. The method was applied successfully to the determination of cefaclor in pharmaceutical formulations.