Artículos de revistas
HPLC determination of nimesulide in tablets by electrochemical detection
Fecha
1998Registro en:
Analytical Letters, Volumen 31, Issue 7, 2018, Pages 1173-1184
00032719
10.1080/00032719808002854
Autor
Álvarez Lueje, Alejandro
Vásquez,
Núñez Vergara, Luis
Squella Serrano, Juan
Institución
Resumen
An analytical chromatographic method for the determination of nimesulide in pharmaceutical forms has been developed. The method is based on high performance liquid chromatographic (HPLC) and electrochemical detection. Chromatography was performed on a μBondapak/μPorasil C-18 column (150 min x 3.9 mm I.D.). The mobile phase consisted of pH 3 buffer phosphate-methanol (40:60, v/v) at a flow-rate of 1 mL/min. The analytes were detected by electrochemical detection in the pulse mode at 1200 mV. For comparative purposes both an HPLC with diode array detector and a spectrophotometric method were employed. The results of the recovery study (mean 101.07, rsd 1.44) show that HPLC with an electrochemical detection method is adequately precise and accurate and can be recommended for the determination of nimesulide in tablets.