Radiation sensor based on MOSFETs mismatch amplification for radiotherapy applications
Fecha
2016-06Registro en:
Garcia Inza, Mariano Andrés; Carbonetto, Sebastián Horacio; Lipovetzky, José; Faigon, Adrian Nestor; Radiation sensor based on MOSFETs mismatch amplification for radiotherapy applications; Institute of Electrical and Electronics Engineers; Ieee Transactions on Nuclear Science; 63; 3; 6-2016; 1784-1789
0018-9499
1558-1578
CONICET Digital
CONICET
Autor
Garcia Inza, Mariano Andrés
Carbonetto, Sebastián Horacio
Lipovetzky, José
Faigon, Adrian Nestor
Resumen
In this paper we present a new dosimeter based on a pair of thick gate oxide MOSFET sensors. A differential circuit topology with a feedback loop provides stabilized output with selectable sensitivity amplification for real time in vivo dosimetry. Radiation response shows a wide linear output range and an effective thermal rejection. These properties make this circuit suitable for dose control in radiotherapy applications.