Artículos de revistas
Conversion coefficients from air kerma to personal dose equivalent in mammography
Fecha
2010Registro en:
RADIATION PHYSICS AND CHEMISTRY, v.79, n.5, p.594-596, 2010
0969-806X
10.1016/j.radphyschem.2009.12.012
Autor
KUNZEL, R.
LEVENHAGEN, R. S.
Herdade, Silvio Bruni
Institución
Resumen
In this work, the scattered X-ray beams produced by a mammography unit with a Mo/Mo, Mo/Rh and W/Rh anode/filter combinations were applied in the evaluation of the Hp(10, 0) and mean conversion coefficients from air kerma to the personal dose equivalent ((C) over barH(p(10,0 degrees))). The higher values of H(p)(10,0 degrees) are related to the Mo/Rh combination whereas the lower ones are for the W/Rh target/filter. (C) over barH(p(10,0 degrees)) values are in the range 0.19-0.54 Sv/Gy, where the higher values comprise the W/Rh combination. (C) 2009 Elsevier Ltd. All rights reserved.