Artículos de revistas
Bedside estimation of recruitable alveolar collapse and hyperdistension by electrical impedance tomography
Fecha
2009Registro en:
INTENSIVE CARE MEDICINE, v.35, n.6, p.1132-1137, 2009
0342-4642
10.1007/s00134-009-1447-y
Autor
COSTA, Eduardo L. V.
BORGES, Joao Batista
MELO, Alexandre
SUAREZ-SIPMANN, Fernando
TOUFEN JR., Carlos
BOHM, Stephan H.
AMATO, Marcelo B. P.
Institución
Resumen
To present a novel algorithm for estimating recruitable alveolar collapse and hyperdistension based on electrical impedance tomography (EIT) during a decremental positive end-expiratory pressure (PEEP) titration. Technical note with illustrative case reports. Respiratory intensive care unit. Patients with acute respiratory distress syndrome. Lung recruitment and PEEP titration maneuver. Simultaneous acquisition of EIT and X-ray computerized tomography (CT) data. We found good agreement (in terms of amount and spatial location) between the collapse estimated by EIT and CT for all levels of PEEP. The optimal PEEP values detected by EIT for patients 1 and 2 (keeping lung collapse < 10%) were 19 and 17 cmH(2)O, respectively. Although pointing to the same non-dependent lung regions, EIT estimates of hyperdistension represent the functional deterioration of lung units, instead of their anatomical changes, and could not be compared directly with static CT estimates for hyperinflation. We described an EIT-based method for estimating recruitable alveolar collapse at the bedside, pointing out its regional distribution. Additionally, we proposed a measure of lung hyperdistension based on regional lung mechanics.