Zone trapping/merging zones in flow analysis: A novel approach for rapid assays involving relatively slow chemical reactions
Fecha
2011Registro en:
TALANTA, v.85, n.1, p.259-263, 2011
0039-9140
10.1016/j.talanta.2011.03.065
Autor
VIDA, Ana C. F.
SASAKI, Milton K.
GOMES, Taciana F.
SILVA, Claudineia R.
FERES, Mario A.
ZAGATTO, E. A. G.
Institución
Resumen
A novel strategy for accomplishing zone trapping in flow analysis is proposed. The sample and the reagent solutions are simultaneously inserted into convergent carrier streams and the established zones merge together before reaching the detector, where the most concentrated portion of the entire sample zone is trapped. The main characteristics, potentialities and limitations of the strategy were critically evaluated in relation to an analogous flow system with zone stopping. When applied to the spectrophotometric determination of nitrite in river waters, the main figures of merit were maintained, exception made for the sampling frequency which was calculated as 189h(-1), about 32% higher relatively to the analogous system with zone stopping. The sample inserted volume can be increased up to 1.0 mL without affecting sampling frequency and no problems with pump heating or malfunctions were noted after 8-h operation of the system. In contrast to zone stopping, only a small portion of the sample zone is halted with zone trapping, leading to these beneficial effects. (C) 2011 Elsevier B.V. All rights reserved.