dc.description.abstractOBJETIVO: o propósito do presente estudo é avaliar o limite de resistência à flexão de um protótipo de mini-implante desenvolvido para ancoragem do aparelho de Herbst. MÉTODOS: após a realização de um cálculo do tamanho da amostra, quatro corpos de prova contendo os protótipos de mini-implantes foram submetidos a uma força de flexão por engastamento simples, utilizando-se uma máquina universal de ensaios mecânicos, sendo calculado o limite de resistência à força de flexão. RESULTADOS: após os ensaios mecânicos, os novos mini-implantes apresentaram o limite de resistência à força de flexão de 98,2kgf, que foi o menor valor encontrado. CONCLUSÃO: os protótipos de mini-implantes desenvolvidos para ancoragem do aparelho de Herbst foram capazes de suportar forças de flexão maiores do que as forças de mordida descritas na literatura.
dc.description.abstractAIM: The purpose of this study was to evaluate the limit of flexural strength of a mini-implant prototype designed for Herbst appliance anchorage. METHODS: After sample size calculation, four specimens with the new mini-implant were submitted to a single cantilever flexure test using a universal testing machine. The limit of flexural force strength was calculated. RESULTS: The mini-implant prototype showed a limit of flexural force strength of 98.2 kgf (982 N), that was the lowest value found. CONCLUSION: The mini-implant prototype designed for Herbst appliance anchorage can withstand flexural forces higher than the maximum human bite forces reported in the literature.
dc.languagepor
dc.publisherDental Press International
dc.relationDental Press Journal of Orthodontics
dc.rightsCopyright Dental Press International
dc.rightsopenAccess
dc.subjectAparelhos ortodônticos
dc.subjectOrtodontia
dc.subjectAparelho de Herbst
dc.subjectMini-implante
dc.subjectOrthodontic appliances
dc.subjectOrthodontics
dc.subjectHerbst appliance
dc.subjectMini-implant
dc.titleAvaliação in vitro da resistência à flexão de um protótipo de mini-implante desenvolvido para ancoragem do aparelho de Herbst