Tesis
Efeito do torque de remoção do implante dental no osso peri-implantar = Effect of removal torque of dental implant in peri-implant bone
Effect of removal torque of dental implant in peri-implant bone
Registro en:
Autor
Silva, Ricardo de Oliveira, 1976-
Institución
Resumen
Orientador: Paulo Henrique Ferreira Caria Tese (doutorado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba Resumo: Objetivo: O objetivo deste estudo foi avaliar o efeito de torque de remoção (de torque reverso) de implantes dentais de titânio removidos do osso peri-implantar, com retriever, de minipigs. Métodos: Os dentes P1-M1 (1º, 2º e 3º pré-molares e 1º molar) foram extraídos bilateralmente de 6 minipigs. Cada animal recebeu 6 implantes de titânio Dentifix® (HE, Ø4.1x10 mm de comprimento, com superfície STA), três para cada lado da mandíbula. De um lado, 3 implantes osseointegraram por 9 meses (9M) com atividade mastigatória fornecida por uma prótese instalada por 3 meses e no lado oposto, outros 3 implantes foram colocados e removidos no mesmo ato cirúrgico (IR). Todos os 36 implantes foram removidos por torque de remoção (retriever) e os valores registrados foram analisados estatisticamente. Os animais foram eutanasiados logo após o torque de remoção. Forma removidos os blocos das áreas peri-implantares e cada terço (cervical, média e apical) do osso peri-implantar foi extraído e analisado histológica e imunoistoquímica. O teste t de Student foi utilizado para determinar diferenças estatísticas nos valores entre as amostras 9M e IR. Os dados foram apresentados como médias e desvios padrão. O nível de significância foi estabelecido em 5% (P <0,05). Resultados: valores de torque de remoção foram maiores na situação experimental 9M do que na IR. Características histológicas do osso maduro foram apresentadas na condição experimental 9M enquanto características ósseas imaturas foram apresentadas na IR. Torque de remoção causou pequenas fraturas e arredondamento no grooving ósseo. A análise imunoistoquímica reforçou os resultados histológicos, o teste t de Student forneceu diferenças estatisticamente significantes para a expressão de osteocalcina em amostras 9M e sem expressão estatisticamente significante para diferenças de colágeno I em ambas as condições experimentais (P <0,05). Conclusões: Remoção de torque causou fraturas microscópicas e suavização nos sulcos ósseos peri-implantares mas não comprometeu a cicatrização óssea Abstract: Purpose: The objective of this study was to evaluate the effect of removal torque (reverse torque) of titanium implants on the peri-implant bone in minipigs. Methods: P1-M1 teeth were extracted bilaterally from six mini pigs (BR-1). Each animal received six titanium implants, three for each side of the mandible Dentifix®(External hexagon, Ø4.1x10 mm, with STA surface). On the right side of the mandible, three implants remained for nine months (9M) under masticatory activity and on the left side the other three implants were placed and immediately removed (IR). All the 36 implants were removed by reverse torque and the recorded values were statistically analyzed. Animals were euthanized right after the removal torque, whose values were recorded. Each third (cervical, medium and apical) of the peri-implant bone was extracted and analyzed histologically and immunohistochemically. Student¿s t-test was used to determine statistical differences in the values between the 9M and IR samples. Data were presented as means with standard deviations. The level of significance was set at 5% (P < 0.05). Results: Removal torque values were higher in 9M than in IR samples in the experimental situation. Histological characteristics of mature bone were presented in the 9M experimental condition, whereas the immature bone characteristics were presented in the IR experimental condition. Removal torque caused small fractures and rounding in the bone grooves. Immunohistochemical analysis reinforced the histological results. The Student¿s t-test showed statistically significant differences in the osteocalcin expression in the 9M samples, whereas no statistically significant differences were found in the expression of collagen I in both experimental conditions (P < 0.05). Conclusions: Removal torque caused microscopic fractures and smoothening of the peri-implant bone grooves, but without compromising the bone healing Doutorado Anatomia Doutor em Biologia Buco-Dental 33003033001P3 CAPES
Ítems relacionados
Mostrando ítems relacionados por Título, autor o materia.
-
Bone level around dental implants adjacent to teeth and implants
Gómez-Guzmán, John Jairo; Arias-Holguín, Yohana Marcela; Pérez, María Isabel; Restrepo-Restrepo, Felipe Augusto; Rodríguez-Medina, Carolina; Botero, Javier Enrique