Evaluation Of Characteristic-to-total Spectrum Ratio: Comparison Between Experimental And A Semi-empirical Model.
Registro en:
Applied Radiation And Isotopes : Including Data, Instrumentation And Methods For Use In Agriculture, Industry And Medicine. , 2015-Jan.
1872-9800
10.1016/j.apradiso.2015.01.011
25600507
Autor
Lopez Gonzales, A H
Tomal, A
Costa, P R
Institución
Resumen
Primary X-ray spectra were measured in the range of 80-150kV in order to validate a computer program based on a semiempirical model. The ratio between the characteristic and total air Kerma was considered to compare computed results and experimental data. Results show that the experimental spectra have higher first HVL and mean energy than the calculated ones. The ratios between the characteristic and total air Kerma for calculated spectra are in good agreement with experimental results for all filtrations used.