Separation of steroid isomers by ion mobility mass spectrometry
Registro en:
Journal Of Chromatography A. Elsevier Science Bv, v. 1310, n. 133, n. 137, 2013.
0021-9673
WOS:000324663300016
10.1016/j.chroma.2013.08.056
Autor
Ahonen, L
Fasciotti, M
af Gennas, GB
Kotiaho, T
Daroda, RJ
Eberlin, M
Kostiainen, R
Institución
Resumen
Ion mobility mass spectrometry performed in a compact traveling wave cell (TWIM-MS) is shown to provide a reliable, fast and repeatable method to separate derivatized steroid isomers. Three steroid isomer pairs were analyzed in their native form and as their p-toluenesulfonyl isocyanate derivatives. The native steroids were separated from each other, but no separation could be attained for the isomers. The derivatized steroid isomers were, however, properly separated by TWIM-MS with peak-to-peak resolutions close to or as high as baseline resolution (Rp-p = 0.77-1.08). (C) 2013 Elsevier B.V. All rights reserved. 1310 133 137 FinPharma Doctoral Program (FPDP) Magnus Ehrnrooth Foundation [ke2011n1] Academy of Finland [251575, 257685] Magnus Ehrnrooth Foundation [ke2011n1] Academy of Finland [251575, 257685]