Detection of periimplant fenestration and dehiscence with the use of two scan modes and the smallest voxel sizes of a cone-beam computed tomography device
Registro en:
Oral Surgery Oral Medicine Oral Pathology Oral Radiology. Elsevier Science Inc, v. 115, n. 1, n. 121, n. 127, 2013.
2212-4403
WOS:000312804800027
10.1016/j.oooo.2012.10.003
Autor
de-Azevedo-Vaz, SL
Vasconcelos, KD
Neves, FS
Melo, SLS
Campos, PSF
Haiter-Neto, F
Institución
Resumen
Objective. To assess the accuracy of cone-beam computed tomography (CBCT) in periimplant fenestration and dehiscence detection, and to determine the effects of 2 voxel sizes and scan modes. Study Design. One hundred titanium implants were placed in bovine ribs in which periimplant fenestration and dehiscence were simulated. CBCT images were acquired with the use of 3 protocols of the i-CAT NG unit: A) 0.2 mm voxel size half-scan (180 degrees); B) 0.2 mm voxel size full-scan (360 degrees); and C) 0.12 mm voxel size full scan (360 degrees). Receiver operating characteristic curves and diagnostic values were obtained. The Az values were compared with the use of analysis of variance. Results. The Az value for dehiscence in protocol A was significantly lower than those of B or C (P < .01). They did not statistically differ for fenestration (P > .05). Conclusions. Protocol B yielded the highest values. The voxel sizes did not affect fenestration and dehiscence detection, and for dehiscence full-scan performed better than half-scan. (Oral Surg Oral Med Oral Pathol Oral Radiol 2013;115:121-127) 115 1 121 127