A high frequency electronic transducer for multiphase flow measurements
Registro en:
Journal of the Brazilian Society of Mechanical Sciences. The Brazilian Society of Mechanical Sciences, v. 21, n. 4, p. 611-621, 1999.
0100-7386
S0100-73861999000400005
10.1590/S0100-73861999000400005
Autor
Belo, Francisco Antônio
Moura, Luiz Felipe Mendes de
Institución
Resumen
This paper describes an electronic transducer for multiphase flow measurement. Its high sensitivity, good signal to noise ratio and accuracy are achieved through an electrical impedance sensor with a special guard technique. The transducer consists of a wide bandwidth and high slew rate differentiator where the lead inductance and stray capacitance effects are compensated. The sensor edge effect is eliminated by using a guard electrode based on the virtual ground potential of the operational amplifier. A theoretical modeling and a calibration method are also presented. The results obtained seem to confirm the validity of the proposed technique. 611 621