Determination of Cr, Fe, Co, Ni, Cu, Zn, As and Pb in liquid chemical waste by energy dispersive X-ray fluorescence
Fecha
2011Registro en:
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, v.287, n.1, p.351-355, 2011
0236-5731
10.1007/s10967-010-0694-2
Autor
ALMEIDA, E.
TAVARES, G. A.
BENDASSOLLI, J. A.
NASCIMENTO FILHO, V. F.
Institución
Resumen
A method for simultaneous determination of Cr, Fe, Co, Ni, Cu, Zn, As e Pb in liquid chemical waste using Energy Dispersive X-Ray Fluorescence (EDXRF) technique was evaluated. A small sample amount (200 mu L) was dried on a 6.35 mu m thickness Mylar film at 60 degrees C and the analyses were carried out using an EDXRF spectrometer operated with an X-ray Mo tube (Zr filter) at 30 kV/20 mA. The acquisition time was 300 s and the Ga element was utilized as internal standard at 25 mg/L for quantitative analysis. The method trueness was assessed by spiking and the detection limit for those elements ranged from 0.39 to 1.7 mg/L. This method is notable because it assists the choice of the more appropriated waste treatment procedure, in which inter elemental interference is a matter of importance. In addition, this inexpensive method allows a non-destructive determination of the elements from (19)K to (92)U simultaneously.